Development of a Miniaturised Implantable Cardiac Monitor with Extended Battery Life

Authors

DOI:

https://doi.org/10.5281/zenodo.22642848

Keywords:

insertable cardiac monitor, implantable loop recorder, device miniaturisation, battery longevity, atrial fibrillation diagnostic yield

Abstract

Background: Insertable cardiac monitors (ICMs) provide continuous, multi-year ambulatory electrocardiographic surveillance for patients with unexplained syncope, cryptogenic stroke, and suspected paroxysmal atrial fibrillation, but realising this capability in a chronically implanted device has required parallel advances in miniaturisation, sensing-electrode design, and battery technology. Objective: This narrative review synthesises published literature on the engineering evolution and clinical performance of miniaturised ICMs, spanning the Medtronic Reveal LINQ, Biotronik BioMonitor, and Abbott Confirm Rx platforms. Methods: We reviewed device-development, engineering, and clinical validation studies describing insertable cardiac monitor volume and sensing-vector architecture, battery longevity and emerging battery-free power strategies, and diagnostic yield data from pivotal trials and real-world registries. Findings: Device volume has fallen severalfold across successive generations, from approximately 9 cc to under 2 cc, while sensing-vector length and R-wave amplitude fidelity have been preserved or improved and projected battery longevity has extended toward four years. Pivotal trial and registry evidence, including CRYSTAL-AF and the SMART and PICTURE registries, demonstrate substantially increased arrhythmia and syncope diagnostic yield with continuous ICM monitoring relative to conventional intermittent assessment. Conclusions: Miniaturisation and battery-technology advances have translated directly into measurable clinical benefit, and continued innovation in low-power circuitry, energy harvesting, and sensing algorithms is likely to further extend the diagnostic reach of insertable cardiac monitoring.

References

Tomson TT, Passman R (2015) The Reveal LINQ insertable cardiac monitor. Expert Review of Medical Devices, 12(1):7-18, DOI:10.1586/17434440.2014.953059.

Sanna T, Diener HC, Passman RS, Di Lazzaro V, Bernstein RA, Morillo CA, Rymer MM, Thijs V, Rogers T, Beckers F, Lindborg K, Brachmann J (2014) Cryptogenic stroke and underlying atrial fibrillation. New England Journal of Medicine, 370(26):2478-2486, DOI:10.1056/NEJMoa1313600.

Pürerfellner H, Sanders P, Pokushalov E, Di Bacco M, Bergemann T, Dekker LRC (2015) Miniaturized Reveal LINQ insertable cardiac monitoring system: first-in-human experience. Heart Rhythm, 12(6):1113-1119, DOI:10.1016/j.hrthm.2015.02.030.

Steffel J, Wright DJ, Schäfer H, Rashid-Fadel T, Lewalter T (2017) Insertion of miniaturized cardiac monitors outside the catheter operating room: experience and practical advice. Europace, 19(10):1624-1629, DOI:10.1093/europace/euw304.

Reinsch N, Ruprecht U, Buchholz J, Diehl RR, Kälsch H, Neven K (2018) The BioMonitor 2 insertable cardiac monitor: Clinical experience with a novel implantable cardiac monitor. Journal of Electrocardiology, 51(5):751-755, DOI:10.1016/j.jelectrocard.2018.05.017.

Piorkowski C, Busch M, Nölker G, Schmitt J, Roithinger FX, Young G, Táborský M, Herrmann G, Schmitz D (2019) Clinical evaluation of a small implantable cardiac monitor with a long sensing vector. Pacing and Clinical Electrophysiology, 42(7):1038-1046, DOI:10.1111/pace.13728.

Reinsch N, Füting A, Höwell D, Neven K (2022) The BIOMONITOR III injectable cardiac monitor: Clinical experience with a novel injectable cardiac monitor. Journal of Clinical Medicine, 11(6):1634, DOI:10.3390/jcm11061634.

Lee JH (2016) Miniaturized human insertable cardiac monitoring system with wireless power transmission technique. Journal of Sensors, 2016:5374574, DOI:10.1155/2016/5374574.

Zhang J, Das R, Zhao J, Mirzai N, Mercer J, Heidari H (2022) Battery-free and wireless technologies for cardiovascular implantable medical devices. Advanced Materials Technologies, 7(6):2101086, DOI:10.1002/admt.202101086.

Duvillier L, Demolder A, Van Renterghem S, De Mey C, West F, De Pooter J (2024) Real-world battery longevity of implantable loop recorders implanted for unexplained syncope: results from a large single-center registry. Journal of Cardiovascular Electrophysiology, 35(11):2134-2143, DOI:10.1111/jce.16420.

Zheng Q, Wei Y, Li D, Zeng Z, Li Z (2025) Choosing energy sources for battery-free pacemakers. Device, 3(6):100746, DOI:10.1016/j.device.2025.100746.

Sanders P, Pürerfellner H, Pokushalov E, Sarkar S, Di Bacco M, Maus B, Dekker LR (2016) Performance of a new atrial fibrillation detection algorithm in a miniaturized insertable cardiac monitor: results from the Reveal LINQ Usability Study. Heart Rhythm, 13(7):1425-1430, DOI:10.1016/j.hrthm.2016.03.005.

Quartieri F, Harish M, Calò L, Ebrahim I, Fusco A, Mester S, Cauti F, Park SJ, Francia P, Giovagnoni M, Adragao P, Vezi B, Lin W, Hutson CS, Grammatico A (2023) New insertable cardiac monitors show high diagnostic yield and good safety profile in real-world clinical practice: results from the international prospective observational SMART Registry. Europace, 25(5):euad068, DOI:10.1093/europace/euad068.

Edvardsson N, Frykman V, van Mechelen R, Mitro P, Mohii-Oskarsson A, Pasquie JL, Ramanna H, Schwertfeger F, Ventura R, Voulgaraki D, Garutti C, Stolt P, Linker NJ, PICTURE Study Investigators (2011) Use of an implantable loop recorder to increase the diagnostic yield in unexplained syncope: results from the PICTURE registry. Europace, 13(2):262-269, DOI:10.1093/europace/euq418.

Lacour P, Dang PL, Huemer M, Parwani AS, Attanasio P, Pieske B, Boldt LH, Haverkamp W, Blaschke F (2017) Performance of the new BioMonitor 2-AF insertable cardiac monitoring system: can better be worse? Pacing and Clinical Electrophysiology, 40(5):516-526, DOI:10.1111/pace.13059.

Swale M, Paul V, Delacroix S, Young G, McSpadden L, Ryu K, Di Fiore D, Santos M, Tan I, Conradie A, Duong M, Schwarz N, Worthley S, Pavia S (2022) Changes in R-wave amplitude at implantation are associated with gender and postures: observations from the Confirm Rx body posture and physical activity study. BMC Cardiovascular Disorders, 22(1):439, DOI:10.1186/s12872-022-02752-0.

Downloads

Published

2026-06-29