Finite Element Analysis of Stress Distribution in a Novel Hip Prosthesis Design
DOI:
https://doi.org/10.5281/zenodo.22485984Keywords:
hip prosthesis, finite element analysis, stress shielding, femoral stem design, biomaterialsAbstract
Background: Aseptic loosening after total hip arthroplasty is closely linked to stress shielding, the unloading of periprosthetic bone caused by stiffness mismatch between metallic femoral stems and host bone. Objective: This narrative review synthesizes published finite element analysis (FEA) and clinical evidence on stress distribution in hip prosthesis stem designs. Methods: Peer-reviewed FEA, biomechanical, and densitometric studies addressing conventional alloys, cross-sectional and fenestration optimization, functionally graded and low-modulus alloys, additively manufactured porous/lattice titanium, and carbon-fiber-reinforced polymer composites were reviewed. Findings: Cobalt-chromium and titanium stems consistently show proximal stress shielding in FEA models; cross-sectional optimization yields modest improvement, while low-modulus graded alloys, porous lattice titanium, and composite stems produce substantially larger, reproducible reductions in strain-energy-based stress-shielding indices, corroborated by patient-specific model validation and longitudinal bone mineral density data. Conclusions: Convergent computational and clinical evidence supports architected porosity, graded stiffness, and composite substitution as the most promising directions for reducing stress shielding, though standardized, validated modeling protocols remain needed to confirm long-term clinical benefit.
References
Sumner DR (2015). Long-term implant fixation and stress-shielding in total hip replacement. Journal of Biomechanics, 48(5), 797-800. DOI: 10.1016/j.jbiomech.2014.12.021.
Huiskes R, Weinans H, van Rietbergen B (1992). The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials. Clinical Orthopaedics and Related Research, 274, 124-134.
Peitgen DS, Innmann MM, Merle C, Gotterbarm T, Moradi B, Streit MR (2018). Periprosthetic Bone Mineral Density Around Uncemented Titanium Stems in the Second and Third Decade After Total Hip Arthroplasty: A DXA Study After 12, 17 and 21 Years. Calcified Tissue International, 103(4), 372-379. DOI: 10.1007/s00223-018-0438-9.
Kayabasi O, Ekici B (2007). The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method. Materials & Design, 28(8), 2269-2277. DOI: 10.1016/j.matdes.2006.08.012. Table 2: Selected finite element analysis studies of hip stem stress distribution referenced in this review. Study Stem Concept Key Reported Finding Evaluated Arabnejad et al. 2017 Fully porous 3D-printed ~70-75% reduction in Ti lattice stem stress-shielding score vs dense Ti-6Al-4V Yamako et al. 2017 Graded low-modulus Reduced Ti-33.6Nb-4Sn stem strain-energy-based stress-shielding index Darwich et al. 2019 Carbon/PEEK and PEEK Reduced proximal stress coated stems shielding vs uncoated metal stem Katz et al. 2018 Patient-specific Subject-specific mapping cemented femur model needed to match measured strains Amjad et al. 2024 Ti-27Nb alloy stem, Adequate safety factors multiple activities across walking, stairs, stumbling loads
Senalp AZ, Kayabasi O, Kurtaran H (2007). Static, dynamic and fatigue behavior of newly designed stem shapes for hip prosthesis using finite element analysis. Materials & Design, 28(5), 1577-1583. DOI: 10.1016/j.matdes.2006.02.015.
Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN (2001). Hip contact forces and gait patterns from routine activities. Journal of Biomechanics, 34(7), 859-871. DOI: 10.1016/S0021-9290(01)00040-9.
Bergmann G, Bender A, Dymke J, Duda G, Damm P (2016). Standardized Loads Acting in Hip Implants. PLOS ONE, 11(5), e0155612. DOI: 10.1371/journal.pone.0155612.
Katz Y, Lubovsky O, Yosibash Z (2018). Patient-specific finite element analysis of femurs with cemented hip implants. Clinical Biomechanics, 58, 74-89. DOI: 10.1016/j.clinbiomech.2018.06.012.
Sabatini AL, Goswami T (2008). Hip implants VII: Finite element analysis and optimization of cross-sections. Materials & Design, 29(7), 1438-1446. DOI:
1016/j.matdes.2007.09.002. 10. Kayabasi O, Erzincanli F (2006). Finite element modelling and analysis of a new cemented hip prosthesis. Advances in Engineering Software, 37(7), 477-483. DOI: 10.1016/j.advengsoft.2005.09.003.
Gross S, Abel EW (2001). A finite element analysis of hollow stemmed hip prostheses as a means of reducing stress shielding of the femur. Journal of Biomechanics, 34(8), 995-1003. DOI: 10.1016/S0021-9290(01)00072-0.
Ikhsan, Prabowo AR, Sohn JM, Triyono J (2020). Finite Element Analysis of Different Artificial Hip Stem Designs Based on Fenestration under Static Loading. Procedia Structural Integrity, 27, 101-108. DOI: 10.1016/j.prostr.2020.07.014.
Yamako G, Janssen D, Hanada S, Anijs T, Ochiai K, Totoribe K, Chosa E, Verdonschot N (2017). Improving stress shielding following total hip arthroplasty by using a femoral stem made of β type Ti-33.6Nb-4Sn with a Young's modulus gradation. Journal of Biomechanics, 63, 135-143. DOI: 10.1016/j.jbiomech.2017.08.017.
Al-Jassir FF, Fouad H, Alothman OY (2013). In vitro assessment of Function Graded (FG) artificial Hip joint stem in terms of bone/cement stresses: 3D Finite Element (FE) study. BioMedical Engineering OnLine, 12(1), 5. 10.1186/1475-925X-12-5.
Chethan KN, Zuber M, Bhat NS, Shenoy BS, Kini CR (2019). Static structural analysis of different stem designs used in total hip arthroplasty using finite element method. Heliyon, 5(6), e01767. DOI: 10.1016/j.heliyon.2019.e01767.
Amjad M, Badshah S, Ahmad S, Badshah M, Jan S, Yasir M, Akram W, Shah IA, Muhammad R, Khan MI, Yasmeen T (2024). Finite element modeling of stress distribution and safety factors in a Ti-27Nb alloy hip implant under real-world physiological loading scenarios. PLOS ONE, 19(8), e0300270. DOI: 10.1371/journal.pone.0300270.
Darwich A, Nazha H, Abbas W (2019). Numerical study of stress shielding evaluation of hip implant stems coated with composite (carbon/PEEK) and polymeric (PEEK) coating materials. Biomedical Research, 30(1). 10.35841/biomedicalresearch.30-18-1048.
Gok MG (2022). Static, fatigue and stress-shielding analysis of the use of different PEEK based materials as hip stem implants. International Polymer Processing, 37(2), 152-163. DOI: 10.1515/ipp-2021-4133.
Ceddia M, Trentadue B (2024). A review of carbon fiber-reinforced polymer composite used to solve stress shielding in total hip replacement. AIMS Materials Science, 11(3), 449-462. DOI: 10.3934/matersci.2024023.
Arabnejad S, Johnston B, Tanzer M, Pasini D (2017). Fully porous 3D printed titanium femoral stem to reduce stress-shielding following total hip arthroplasty. Journal of Orthopaedic Research, 35(8), 1774-1783. 10.1002/jor.23445.
Zhao L, Wang Y, Wang Q, Zhang Y, Yang G (2024). Optimization Design and SLM Manufacturing of Porous Titanium Alloy Femoral Stem. Materials, 17(19), 4896. 10.3390/ma17194896.
Tyagi SA, Manjaiah M (2023). Additive manufacturing of titanium-based lattice structures for medical applications – A review. Bioprinting, 30, e00267. 10.1016/j.bprint.2023.e00267.
Distefano F, Pasta S, Epasto G (2023). Titanium Lattice Structures Produced via Additive Manufacturing for a Bone Scaffold: A Review. Journal of Functional Biomaterials, 14(3), 125. DOI: 10.3390/jfb14030125.
Abu Qudeiri JE, Abdudeen A, Sahadevan MR, Padmanabhan MA (2024). Numerical investigation on the characteristics of hip implant under static loading. Heliyon, 10(4), e26151. DOI: 10.1016/j.heliyon.2024.e26151.
Babić M, Verić O, Božić Ž, Sušić A (2020). Finite element modelling and fatigue life assessment of a cemented total hip prosthesis based on 3D scanning. Engineering Failure Analysis, 113, 104536. DOI: 10.1016/j.engfailanal.2020.104536.
Bliss O, Swadener JG, Pierce G, Zidane I (2023). Finite element analysis of a stemmed hip prosthesis to reduce stress shielding in the proximal femur. Journal of Mechanical Engineering and Sciences, 17(1), 9349-9359. 10.15282/jmes.17.1.2023.5.0739. DOI: DOI: DOI: DOI: DOI: wear DOI: